Skip to main content

Research Repository

Advanced Search

Biomechanical analysis of tension band wiring (TBW) of transverse fractures of patella

Kuiper

Biomechanical analysis of tension band wiring (TBW) of transverse fractures of patella Thumbnail


Authors



Abstract

Abstract
Purpose
Tension band wiring is commonly used for fixation of simple transverse fractures. The popular configuration is parallel Kirschner wires (K-wires) and a stainless steel wire loop placed in a vertically oriented figure-of-8.

Methods
We used a wooden model of a patella with a midway transverse fracture and compared four different types of fixation. The first construct had a vertical figure-of-8 with one twist of wire. The second contained a vertical figure-of-8 with two twists of wire. The third was a vertical figure-of-8 with two twists of wire placed at adjacent corners while the last one had a horizontal figure-of-8 with two twists of wire placed at adjacent corners. Interfragmentary compression at the point of wire breakage was measured for each construct as well as permanent displacement on cyclic loading.

Results
Placement of the figure-of-eight in a horizontal orientation with two wire twists at the corner improved interfragmentary compression by 63% (p < 0.05, Tukey post hoc test). On cyclic loading, all the constructs with vertical figure-of-eight but none with a horizontal construct failed (p = 0.01; Fisher's exact test). Permanent fracture displacement after cyclic loading was 67% lower with horizontal figure-of-eight constructs (p < 0.05; t test).

Conclusion
Placing wire twists at the corner and a horizontal placement of figure-of-8 improves stability of the construct.

Acceptance Date Jun 25, 2016
Publication Date Sep 6, 2016
Journal Chinese Journal of Traumatology
Print ISSN 1008-1275
Publisher Elsevier
Pages 255-258
DOI https://doi.org/10.1016/j.cjtee.2016.06.009
Keywords Patella transverse fractures, Tension band wiring, Biomechanical analysis
Publisher URL http://dx.doi.org/10.1016/j.cjtee.2016.06.009