Introduction
40 mm large diameter heads offer the advantages of lesser dislocation rates and better stability while highly cross linked polyethylene have lower wear rates than ultra high molecular weight polyethylene. Studies of the survivorship of 40 mm heads in hybrid hip replacements with Exeter stem and second generation highly cross linked polyethylene are limited. The purpose of the study is to report the short term of survivorship of the large diameter heads (40 mm) with Exeter stem with the secondary aim being the survival analysis of the thinnest second generation highly cross linked polyethylene.

Methods
Retrospective case series of survivorship of patients with hybrid hip replacements of Exeter stems with 40 mm heads articulating with second generation triple annealed highly cross linked polyethylene liner on a uncemented acetabular shell was performed. As a subset, survival of thinnest second generation highly cross linked polyethylene survival (3.8 mm) at short term was assessed. Survival of the implants was confirmed from the hospital records and National joint registry as of 2015. Revision for any cause was taken as end point.

Results
324 hybrid hip replacements with 40 mm heads had been performed for primary hip osteoarthritis. Of the 324 hip replacements, 154 hip replacements had thinnest second generation highly cross linked polyethylene (3.8 mm). Two patients had revision of components, one for periprosthetic fracture and one for deep infection. Mean age of the patients was 70.5 years (range 42–88 years, median 71, SD 8.3 years). None of the patients had revision due to trunion wear or loosening of components. The overall 5-year implant survival probability of hips with 40 mm heads was 99.4% (95% CI 98 to 100%) while the subset group of hip replacements with thinnest second generation highly cross linked polyethylene (3.8 mm) had 5-year implant survival probability of 99.3% (95% CI 97.1 to 100%).

Conclusion
Short term survivorship does not show significant evidence of early failure or higher rate of revision in our series of hybrid hip replacements with large diameter heads and second generation triple annealed highly cross linked polyethylene. Dislocation rate at the short term is none. Results from this series have to be carefully interpreted due to the relatively short follow up but so far results are encouraging. Long term follow up is required to conclude whether there is early or higher rate of failure. It is our intention to follow up this cohort and further publish our results at longer term.

Item Type: Article
Additional Information: This is the accepted author manuscript (AAM). The final published version (version of record) is available online via Elsevier at https://doi.org/10.1016/j.jcot.2017.12.008 Please refer to any applicable terms of use of the publisher.
Uncontrolled Keywords: large diameter heads; highly cross linked polyethylene; sequentially processed polyethylene; second generation HCLPE; revision; low failure
Subjects: R Medicine > R Medicine (General)
R Medicine > RD Surgery > RD32 Operative surgery. Technique of surgical operations