Keele Research Repository
Explore the Repository
Menown, IBA, Mamas, MA, Cotton, JM, Hildick-Smith, D, Eberli, FR, Leibundgut, G, Tresukosol, D, Macaya, C, Copt, S, Slama, SS and Oldroyd, KG (2021) Thin Strut CoCr Biodegradable Polymer Biolimus A9-Eluting Stents versus Thicker Strut Stainless Steel Biodegradable Polymer Biolimus A9-Eluting Stents: Two-Year Clinical Outcomes. Journal of Interventional Cardiology, 2021. ISSN 0896-4327
Thin Strut CoCr Biodegradable Polymer Biolimus A9-Eluting Stents versus Thicker Strut Stainless Steel Biodegradable Polymer .pdf - Published Version
Download (432kB) | Preview
Abstract
Background: While thinner struts are associated with improved clinical outcomes in bare-metal stents (BMS), reducing strut thickness may affect drug delivery from drug-eluting stents (DES) and there are limited data comparing otherwise similar thin and thick strut DES. We assessed 2-year outcomes of patients treated with a thin strut (84-88um) cobalt-chromium, biodegradable polymer, Biolimus A9-eluting stent (CoCr-BP-BES) and compared these to patients treated with a stainless steel, biodegradable polymer, Biolimus A9-eluting stent (SS-BP-BES).
Methods: In total, 1257 patients were studied: 400 patients from 12 centres receiving ≥1 CoCr-BP-BES in the prospective Biomatrix Alpha registry underwent prespecified comparison with 857 patients who received ≥1 Biomatrix Flex SS-BP-BES in the LEADERS study (historical control). The primary outcome was major adverse cardiac events (MACE)-cardiac death, myocardial infarction (MI), or clinically driven target vessel revascularization (cd-TVR). Propensity analysis was used to adjust for differences in baseline variables and a landmark analysis at day-3 to account for differences in periprocedural MI definitions.
Results: MACE at 2 years occurred in 6.65% CoCr-BP-BES versus 13.23% SS-BP-BES groups (unadjusted HR 0.48 [0.31-0.73]; P=0.0005). Following propensity analysis, 2-year adjusted MACE rates were 7.4% versus 13.3% (HR 0.53 [0.35-0.79]; P=0.004). Definite or probable stent thrombosis, adjudicated using identical criteria in both studies, occurred less frequently with CoCr-BP-BES (1.12% vs. 3.22%; adjusted HR 0.32 [0.11-0.9]; P=0.034). In day-3 landmark analysis, the difference in 2-year MACE was no longer significant but there was a lower patient-orientated composite endpoint (11.7% vs. 18.4%; HR 0.6 [0.43-0.83]; P=0.006) and a trend to lower target vessel failure (5.8% vs. 9.1%; HR 0.63 [0.4-1.00]; P=0.078).
Conclusion: At 2-year follow-up, propensity-adjusted analysis showed the thin strut (84-88um) Biomatrix Alpha CoCr-BP-BES was associated with improved clinical outcomes compared with the thicker strut (114-120um) Biomatrix Flex SS-BP-BES.
Item Type: | Article |
---|---|
Additional Information: | Copyright © 2021 Ian B. A. Menown et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
Subjects: | R Medicine > R Medicine (General) > R735 Medical education. Medical schools. Research R Medicine > RD Surgery |
Divisions: | Faculty of Medicine and Health Sciences > School of Primary, Community and Social Care |